MAUDE Adverse Event Report: SISTEMAS MEDICOS ALARIS, S.A. DE C.V. BD MAXZERO¿ MULTI-FUSE PRESSURE RATED EXTENSION SET WITH NEEDLELESS CONNECTOR; INTRAVASCULAR ADMINISTRATION SET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SISTEMAS MEDICOS ALARIS, S.A. DE C.V. BD MAXZERO¿ MULTI-FUSE PRESSURE RATED EXTENSION SET WITH NEEDLELESS CONNECTOR; INTRAVASCULAR ADMINISTRATION SET Back to Search Results
Catalog Number MZ5309
Device Problems Leak/Splash (1354); Loose or Intermittent Connection (1371)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/03/2023
Event Type  malfunction  
Manufacturer Narrative
Investigation summary: a complaint of a set not connecting properly was received from the customer.No product or photo was returned by the customer.The customer complaint of connection issues could not be verified due to the product not being returned for failure investigation.A device history record review for model mz5309 lot number 22099132 was performed.The search showed that a total of (b)(4) in 1 lot number were built on 07sep2022.There were no quality notifications issued for the failure mode reported by the customer during the production build of this set.Due to no sample being received, an investigation could not be performed, and a root cause could not be determined.This incident has been added to our database of reported incidents.Our business team regularly reviews the collected data for identification of emerging trends.
 
Event Description
It was reported that the bd maxzero¿ multi-fuse pressure rated extension set with needleless connector had a loose connection to the catheter and would not properly tighten, causing blood to leak out.The following information was provided by the initial reporter: "extension tubing that screws into catheter after an iv start is not twisting properly.It doesn't spin so it will not tighten down and blood gets everywhere.The cap is too stiff and difficult to maneuver while starting the iv.I've tried loosening it prior to beginning the iv process but if the cap is crooked at all it still does it.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD MAXZERO¿ MULTI-FUSE PRESSURE RATED EXTENSION SET WITH NEEDLELESS CONNECTOR
Type of Device
INTRAVASCULAR ADMINISTRATION SET
Manufacturer (Section D)
SISTEMAS MEDICOS ALARIS, S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer (Section G)
SISTEMAS MEDICOS ALARIS, S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8448235433
MDR Report Key16475652
MDR Text Key310592579
Report Number9616066-2023-00310
Device Sequence Number1
Product Code FPA
UDI-Device Identifier10885403242083
UDI-Public10885403242083
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K140831
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 02/08/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue NumberMZ5309
Device Lot Number22099132
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/03/2023
Initial Date FDA Received03/02/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/07/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age2 MO
-
-