There was no patient involvement.Livanova deutschland manufactures the heater-cooler system 3t.The incident occurred in los angeles, california.A livanova field service representative was dispatched to the facility to investigate the device and could reproduce the reported issue.Initially, central processing unit (cpu) board was replaced without solve the issue.Then, temperature probes were checked: cardioplegia side was reading properly, while patient side was reading off one (1) degree (lower than system reading).In order to fix the issue, temperature probes of the patient tank were re-calibrated.Subsequent functional verification testing was completed without further issues and the unit was returned to service.Most likely root cause of the reported event was traced back to a temperature probe of the patient tank with a calibration that drifted over time and was corrected by livanova technician during troubleshooting.
|