• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER CENTURION VISION SYSTEM, ACCESSORY, OZIL TORSIONAL HANDPIECE; UNIT, PHACOFRAGMENTATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER CENTURION VISION SYSTEM, ACCESSORY, OZIL TORSIONAL HANDPIECE; UNIT, PHACOFRAGMENTATION Back to Search Results
Catalog Number 8065751761
Device Problems Particulates (1451); Excessive Heating (4030)
Patient Problems Blurred Vision (2137); Discomfort (2330); Eye Burn (2523)
Event Date 09/22/2022
Event Type  Injury  
Manufacturer Narrative
Investigation, including root cause analysis, is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event Description
A customer reported the patient underwent cataract surgery in the right eye.During the cataract surgery from an ophthalmic operating handpiece and phacoemulsification tip a milky white substance emerged and occluded.While using an ophthalmic handpiece which was hot and it was resulted in the wound burn, vision was blurry and minor discomfort.Burn was sutured.Currently patient was recovered.
 
Manufacturer Narrative
The product under investigation is not a serviceable device.Therefore, a service record review was not performed.However, there was no product returned for this investigation.Based on the information obtained, the root cause of the reported event is inconclusive.Specific product identifiers (serial number) were not provided and could not be determined at this time.However, all device history records are reviewed prior to product release to ensure the product was manufactured in compliance with the device master record and meets release criteria.A review for complaints reported against this serial number cannot be performed as the serial number is unknown.The serial is unknown.Therefore, a service history review cannot be performed.Corneal burn is an issue that is occasionally reported with cataract surgery.According to the pennsylvania patient safety advisory abstract: preventing corneal burns during phacoemulsification, march 2010, vol.7, no.1: 23-25, most corneal burns can be traced to issues related to surgical technique and not to malfunctioning equipment.Based on the information obtained, the root cause of the reported event is inconclusive.Manufacturer will continue to monitor data for evidence of adverse trending and take further action, as appropriate.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTURION VISION SYSTEM, ACCESSORY, OZIL TORSIONAL HANDPIECE
Type of Device
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer (Section G)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key16477789
MDR Text Key310600098
Report Number2028159-2023-00277
Device Sequence Number1
Product Code HQC
UDI-Device Identifier00380657517619
UDI-Public00380657517619
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K121555
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 06/09/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue Number8065751761
Device Lot NumberASKU
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 02/03/2023
Initial Date FDA Received03/03/2023
Supplement Dates Manufacturer Received03/22/2023
Supplement Dates FDA Received06/09/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
DUREZOL QID; KETORALAC QID; OFLOXACIN QID; PHACO TIP
Patient Outcome(s) Other; Required Intervention;
-
-