Model Number 2404-03 |
Device Problems
Entrapment of Device (1212); Material Deformation (2976)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/09/2023 |
Event Type
Injury
|
Event Description
|
It was reported that device entrapment occurred.The 90% stenosed target lesion was located in the mildly tortuous and severely calcified atrioventricular branch.A fractional flow reserve procedure was performed using a comet ii pressure guidewire.During the procedure, the patient sneezed causing the impermeable part of the tip to be trapped in a small peripheral blood vessel.The detached tip was successfully retrieved by using a microcatheter.The wire was deformed and stretched when it was pulled out.The procedure was completed with another of the same device.No patient complications were reported, and the patient had no problems following the procedure.
|
|
Manufacturer Narrative
|
Device evaluated by mfr.: the device was returned for analysis.Returned product consisted of an ffr comet pressure wire.The occ cable was not returned for analysis.The tip and device shaft were visually and microscopically examined for damage or any irregularities.Inspection of the device revealed that the tip was bent and stretched 6.5cm in length.The tip was also partially separated/detached approximately 2.5cm proximal of the proximal end of the tip (no pieces were missing as the separated ends were still intact to the stretched wire) due to the stretching.The appearance of the confirmed tip damage is consistent to damage that can be caused from interaction with another device or patient anatomy during the procedure.
|
|
Event Description
|
It was reported that device entrapment occurred.The 90% stenosed target lesion was located in the mildly tortuous and severely calcified atrioventricular branch.A fractional flow reserve procedure was performed using a comet ii pressure guidewire.During the procedure, the patient sneezed causing the impermeable part of the tip to be trapped in a small peripheral blood vessel.The detached tip was successfully retrieved by using a microcatheter.The wire was deformed and stretched when it was pulled out.The procedure was completed with another of the same device.No patient complications were reported, and the patient had no problems following the procedure.
|
|
Search Alerts/Recalls
|