SHIRAKAWA OLYMPUS CO., LTD. EVIS EXERA III XENON LIGHT SOURCE Back to Search Results
Model Number CLV-190
Device Problem Device Alarm System (1012)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/17/2023
Event Type  malfunction  
Event Description
The customer reported that his olympus evis exera iii xenon light source¿s front panel leds were all flashing during procedure preparation.According to the initial reporter the issue was resolved by checking/cleaning one of the switches in the connector.There was no patient harm reported as a result of this event.
 
Manufacturer Narrative
The device has not been returned for evaluation.If additional information becomes available prior to the conclusion of the investigation, a supplemental report will be filed.
 
Manufacturer Narrative
This report is being supplemented to provide results of the investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.There was no indication that the event was caused by a misuse or that the event was related to design of the device.Repair history was reviewed and no issues were found related to the reported event.During troubleshooting, checking and cleaning one of the switches on the connector resolved the issue of the front panel blinking.Although it was determined that defect was likely due to the connector's switch, a definitive root cause could not be identified.If additional information becomes available at a later date, this report will be supplemented.Olympus will continue to monitor the field performance of this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA III XENON LIGHT SOURCE
Type of Device
XENON LIGHT SOURCE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key16478123
MDR Text Key310858061
Report Number3002808148-2023-02036
Device Sequence Number1
Product Code NWB
UDI-Device Identifier04953170298868
UDI-Public04953170298868
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K131780
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 04/25/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCLV-190
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 02/17/2023
Initial Date FDA Received03/03/2023
Supplement Dates Manufacturer Received03/31/2023
Supplement Dates FDA Received04/25/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-