Catalog Number CV-17702-E
Device Problem Unraveled Material (1664)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/28/2022
Event Type  malfunction  
Manufacturer Narrative
Qn# (b)(4).
 
Event Description
The complaint is reported as: "upon removal of the guidewire, damage to it is evidenced." no patient harm was reported.Another device was successfully used.The patient's condition is reported as fine.
 
Event Description
The complaint is reported as: "upon removal of the guidewire, damage to it is evidenced." no patient harm was reported.Another devcie was successfully used.The patient's condition is reported as fine.
 
Manufacturer Narrative
(b)(4).The actual device was not returned; however, the customer provided one photo for analysis.The complaint of spring wire guide (swg) unraveled in use was able to be confirmed by the photo.The image shows an unraveled guide wire with evidence of use (dried blood).A complete visual inspection could not be performed as no sample was returned for analysis.A device history record review was performed, and no relevant findings were identified.The instructions for use (ifu) provided with this kit warns the user, "do not apply excessive force in placing or removing catheter or guidewire.Excessive force can cause component damage or breakage.If damage is suspected or withdrawal cannot be easily accomplished, radiographic visualization should be obtained and further consultation requested.Without the device to evaluate, the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARROW CVC SET: 2-LUMEN 7 FR X 20 CM
Type of Device
CATHETER PERCUTANEOUS
Manufacturer (Section D)
ARROW INTERNATIONAL LLC
morrisville NC
Manufacturer (Section G)
ARROW INTERNATIONAL DE MEXICO S.A. DE C.V.
ave. washington 3701
colonia panamericana, chihuahua
chihuahua 31200
MX   31200
Manufacturer Contact
katharine tarpley
3015 carrington mill blvd
morrisville, NC 27560
MDR Report Key16478585
MDR Text Key310806855
Report Number9680794-2023-00124
Device Sequence Number1
Product Code DQY
Combination Product (y/n)N
Reporter Country CodeCO
PMA/PMN Number
K862056
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative,Distributor
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 02/03/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberCV-17702-E
Device Lot Number14F22A0083
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 02/03/2023
Initial Date FDA Received03/03/2023
Supplement Dates Manufacturer Received03/21/2023
Supplement Dates FDA Received03/28/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/24/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial