Catalog Number 5003415 |
Device Problems
Off-Label Use (1494); Use of Device Problem (1670)
|
Patient Problems
Discomfort (2330); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
Injury
|
Manufacturer Narrative
|
This incident is reportable according to 21 cfr 803.The fda defines this as a serious injury (21 cfr sec.803.3) the incident will be reported to maintain compliance with 21 cfr 803 and out of an abundance of caution.An infants father notified kulzer that in (b)(6) 2022 his son had an impression taken with xantopren comfort medium for cleft palate surgery.Some of the material was swallowed and ingested.The infant extruded (bowel movement) some of the material 4 days after and then again in (b)(6) 2023, according to the father.The ifu contains the following warning and precaution: "do not swallow or ingest.If health problems arise after swallowing impression material, seek medical attention immediately.Intestinal blockage may arise in rare cases.".As the baby was 3 months old when the impression was performed, the dentist did not follow the additional warning and precaution: "do not use in children under 6 years of age." this product is not sold in the usa or canada, however, a comparable product is.We will report this incident out of an abundance of caution.
|
|
Event Description
|
An infants father notified kulzer that in (b)(6) 2022 his son had an impression taken with xantopren comfort medium for cleft palate surgery.Some of the material was swallowed and ingested.The infant extruded (bowel movement) some of the material 4 days after and then again in (b)(6) 2023, according to the father.
|
|
Event Description
|
An infants father notified kulzer that in (b)(6) 2022 his son had an impression taken with xantopren comfort medium for cleft palate surgery.Some of the material was swallowed and ingested.The infant extruded (bowel movement) some of the material 4 days after and then again in (b)(6) 2023, according to the father.
|
|
Manufacturer Narrative
|
This incident is reportable according to 21 cfr 803.The fda defines this as a serious injury (21 cfr sec.803.3) the incident will be reported to maintain compliance with 21 cfr 803 and out of an abundance of caution.An infants father notified kulzer that in (b)(6) 2022 his son had an impression taken with xantopren comfort medium for cleft palate surgery.Some of the material was swallowed and ingested.The infant extruded (bowel movement) some of the material 4 days after and then again in (b)(6) 2023, according to the father.The ifu contains the following warning and precaution: "do not swallow or ingest.If health problems arise after swallowing impression material, seek medical attention immediately.Intestinal blockage may arise in rare cases." as the baby was 3 months old when the impression was performed, the dentist did not follow the additional warning and precaution: "do not use in children under 6 years of age." this product is not sold in the usa or canada, however, a comparable product is.We will report this incident out of an abundance of caution.
|
|
Search Alerts/Recalls
|