Model Number CLV-190
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
The customer reported to olympus that the evis exera iii xenon light source always switches to emergency mode.There were no reports of patient harm associated with this event.During the evaluation of the device, it was noted that due to poor contact with the output socket, the scope detection function did not work.This report is to capture the reportable malfunction of scope detection not working noted at estimation.
 
Manufacturer Narrative
The device was evaluated by olympus.The evaluation found that the allegation was not confirmed.Additional issues were identified during the device evaluation: a 3rd party lamp was installed; the pump loses pressure quickly; due to damage on the power switch, the switch was stuck and could not be turned off; due to damage on the pump, the airflow was too weak.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 10 years since the subject device was manufactured.Based on the results of the investigation, the light source switching to the emergency lighting was not confirmed during evaluation, the definitive root cause of the reported issue could not be determined.A probable cause was that it was on because a third-party lamp was used.Based on the results of the investigation, it is likely the scope not being detected occurred due to poor contact of the output socket.However, a definitive root cause could not be determined.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA III XENON LIGHT SOURCE
Type of Device
XENON LIGHT SOURCE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key16486401
MDR Text Key311481249
Report Number3002808148-2023-02131
Device Sequence Number1
Product Code NWB
UDI-Device Identifier04953170298868
UDI-Public04953170298868
Combination Product (y/n)N
Reporter Country CodeAU
PMA/PMN Number
K131780
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 04/25/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCLV-190
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/13/2023
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 02/07/2023
Initial Date FDA Received03/06/2023
Supplement Dates Manufacturer Received03/29/2023
Supplement Dates FDA Received04/25/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/04/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-