Model Number X SERIES |
Device Problems
Erratic or Intermittent Display (1182); No Display/Image (1183)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
|
|
Event Description
|
Complainant alleged that during a routine shift check by a clinician, the device's display showed a blue line down the screen and was not legible.Complainant indicated that there was no patient involvement in the reported malfunction.
|
|
Event Description
|
Complainant alleged that during a routine shift check by a clinician, the device's display showed a blue line down the screen, but the screen was still legible.Complainant indicated that there was no patient involvement in the reported malfunction.
|
|
Manufacturer Narrative
|
This supplemental medwatch report is reporting the evaluation of the device.This supplemental medwatch report is also correcting information submitted on the initial medwatch report.Please reference sections b5 and h6:device problem code.Evaluation results: the device was returned to zoll medical corporation for evaluation.The customer's report was observed and attributed to the lcd display.The lcd display was replaced to resolve the report.Analysis of reports of this type has not identified an increase in trend.Reports of this nature are typically not considered to be medwatch reportable as there is no potential for clinical impact.The screen and critical information was still visibale to the user.
|
|