HEARTSINE TECHNOLOGIES LTD PACKAGE, 350P, PP03, ES, 350-STR-ES-10; AUTOMATED EXTERNAL DEFIBRILLATOR
|
Back to Search Results |
|
Model Number 350P |
Device Problem
Electrical /Electronic Property Problem (1198)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/27/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
Heartsine's investigation of the device confirmed the reported fault as upon return, a faulty pogo pin was identified.This had resulted in the voltage fluctuations and low battery fails.The fault was attributed to a faulty pogo pin.The device was scrapped by heartsine and the customer was provided with a replacement device.
|
|
Event Description
|
The distributor contacted heartsine to report a non-critical issue with a customer's device.Upon evaluation of the customer¿s device, heartsine a faulty pogo pin.This may prevent the device from powering on which may prevent or delay defibrillation therapy.There was no patient involvement reported with the event.
|
|