MAUDE Adverse Event Report: HEARTSINE TECHNOLOGIES LTD PACKAGE, 350P, PP03, ES, 350-STR-ES-10; AUTOMATED EXTERNAL DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HEARTSINE TECHNOLOGIES LTD PACKAGE, 350P, PP03, ES, 350-STR-ES-10; AUTOMATED EXTERNAL DEFIBRILLATOR Back to Search Results
Model Number 350P
Device Problem Electrical /Electronic Property Problem (1198)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/27/2022
Event Type  malfunction  
Manufacturer Narrative
Heartsine's investigation of the device confirmed the reported fault as upon return, a faulty pogo pin was identified.This had resulted in the voltage fluctuations and low battery fails.The fault was attributed to a faulty pogo pin.The device was scrapped by heartsine and the customer was provided with a replacement device.
 
Event Description
The distributor contacted heartsine to report a non-critical issue with a customer's device.Upon evaluation of the customer¿s device, heartsine a faulty pogo pin.This may prevent the device from powering on which may prevent or delay defibrillation therapy.There was no patient involvement reported with the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PACKAGE, 350P, PP03, ES, 350-STR-ES-10
Type of Device
AUTOMATED EXTERNAL DEFIBRILLATOR
Manufacturer (Section D)
HEARTSINE TECHNOLOGIES LTD
207 airport road west
belfast BT3 9 ED
EI  BT3 9ED
Manufacturer (Section G)
HEARTSINE TECHNOLOGIES LTD
207 airport road west
belfast BT3 9 ED
EI   BT3 9ED
Manufacturer Contact
todd bandy
207 airport road west
belfast BT3 9-ED
EI   BT3 9ED
4258674577
MDR Report Key16493035
MDR Text Key310794072
Report Number3004123209-2023-00021
Device Sequence Number1
Product Code NSA
Combination Product (y/n)N
Reporter Country CodeSP
PMA/PMN Number
P160008
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 03/06/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number350P
Device Catalogue Number350-STR-ES-10
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/11/2023
Initial Date Manufacturer Received 02/13/2023
Initial Date FDA Received03/06/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/22/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-