The device was returned and evaluated by olympus.In addition to the findings, evaluation found the complaint was not confirmed.Evaluation did find the distal end was burned, the bending section cover adhesive was chipped, the bending angle in the up direction did not meet the standard value due to wear of the angle wire, the image was stained, due to damage on the image guide bundle, and multiple parts of the scope were scratched.This event is under investigation.A supplemental report will be submitted upon receiving additional information.
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the root cause of the reported event is unable to be determined.However, the cause of the event is likely due to stress of repeated use, external factors or handling of the device.The event can be detected by following the instructions for use (ifu) which state: ¿ chapter 3 preparation and inspection, 3.3 inspection of the endoscope¿ describes the following.[inspection of the endoscope] 1.Inspect the eyepiece section and appearance of the control section for excessive scratching, deformation, loose parts, or other irregularities visually and with touching.2 inspect the eyepiece lens and the cover glass of the light guide connector for any irregularities such as scratches, cracks, and stains.3 inspect the boot and the insertion section near the boot for any irregularities such as bends, twists, tears, and cracks.4 inspect the external surface of the entire insertion section including the bending section and the distal end for any irregularities such as dents, bulges, swelling, scratches, peeling of coating, holes, sagging, transformation, bends, adhesion of foreign bodies, missing parts, and protruding objects.5.Holding the control section with one hand, carefully run your other hand back and forth over the entire length of the insertion section.Confirm that no objects or metallic wire protrude from the insertion section.Also, confirm that the insertion tube is not abnormally rigid.Olympus will continue to monitor field performance for this device.
|