DATASCOPE CORP. - MAHWAH CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 110V; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
|
Back to Search Results |
|
Model Number 0998-00-3023-53 |
Device Problems
Electrical /Electronic Property Problem (1198); Device Displays Incorrect Message (2591)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/27/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that the cs300 intra-aortic balloon pump (iabp) unit displayed an electrical test fails code 52 while the customer attempted to start up device.There was no patient involved reported.
|
|
Manufacturer Narrative
|
A supplemental report will be submitted upon completion of our investigation.
|
|
Manufacturer Narrative
|
Corrected fields: d5, e3, h6 (medical device ¿ problem code).A getinge field service engineer (fse) evaluated the unit and found that unit had electrical fail fault #52 in logs.History shows drive manifold and drive transducer were just replaced.To fix this issue fse replaced front end board.Unit passed complete pm with full calibration, functional, and electrical safety tests to factory specifications.Unit is cleared for clinical use and returned to customer.
|
|