Model Number 0998-00-3023-53
Device Problems Electrical /Electronic Property Problem (1198); Device Displays Incorrect Message (2591)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/27/2023
Event Type  malfunction  
Event Description
It was reported that the cs300 intra-aortic balloon pump (iabp) unit displayed an electrical test fails code 52 while the customer attempted to start up device.There was no patient involved reported.
 
Manufacturer Narrative
A supplemental report will be submitted upon completion of our investigation.
 
Manufacturer Narrative
Corrected fields: d5, e3, h6 (medical device ¿ problem code).A getinge field service engineer (fse) evaluated the unit and found that unit had electrical fail fault #52 in logs.History shows drive manifold and drive transducer were just replaced.To fix this issue fse replaced front end board.Unit passed complete pm with full calibration, functional, and electrical safety tests to factory specifications.Unit is cleared for clinical use and returned to customer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 110V
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key16496107
MDR Text Key310812963
Report Number2249723-2023-01402
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567107882
UDI-Public10607567107882
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K063525
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 03/08/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number0998-00-3023-53
Device Catalogue Number0998-00-3023-53
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/27/2023
Initial Date FDA Received03/07/2023
Supplement Dates Manufacturer Received03/07/2024
Supplement Dates FDA Received03/08/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/30/2011
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-