Catalog Number 107493
Device Problems Device Emits Odor (1425); Smoking (1585)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/07/2023
Event Type  malfunction  
Event Description
It was reported that during continuous renal replacement therapy (crrt) using a prismaflex system 3.20, smoke was observed coming out from the machine due to a defective power supply (not further specified).The patient was disconnected from the machine without returning the extracorporeal (ec) blood to the patient.The patient was taken out of the room.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
A service history review revealed no indication that the parts replaced during servicing caused or contributed to the reported event.The device was not received for evaluation; therefore, a device analysis could not be completed.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRISMAFLEX MACHINES
Type of Device
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - LUND MONITORS
magistratsvagen 16
lund skane lan SE-22 643
SW   SE-22643
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key16502401
MDR Text Key310951012
Report Number9616026-2023-00013
Device Sequence Number1
Product Code KDI
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K072093
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 03/08/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number107493
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/07/2023
Initial Date FDA Received03/08/2023
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-