• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RAYNHAM CODMAN DISPOS PERFORATOR; DISPOSABLE PERFORATORS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RAYNHAM CODMAN DISPOS PERFORATOR; DISPOSABLE PERFORATORS Back to Search Results
Model Number 261221
Device Problem Insufficient Information (3190)
Patient Problem Insufficient Information (4580)
Event Date 02/02/2023
Event Type  malfunction  
Manufacturer Narrative
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
 
Event Description
A physician reported a disposable perforator (id 261221) was damaged when trying to remove the tip from the burr hole after drilling.According to information provided, the manufacturer of the drill is unknown, it is also unknown if the drill was electric or pneumatic, or if the perforator clicked in place on the drill. it is also unknown if the recommended spring tests were performed between each burr hole.
 
Manufacturer Narrative
The perforator (id 261221) has been returned for evaluation.Device history record (dhr)- there is no indication that the production process may have contributed to this complaint.All test results passed procedural specifications.Failure analysis - visual inspection utilizing the unaided eye performed.The unit was received disassembled and was soiled with a worn eto label.Spring test could not be performed "as is" due to being received disassembled.Functional test could not be performed due to the unit being received disassembled.Root cause - per the failure analyst ¿unit had a shallow weld this weld likely contributed to the disassembly of the unit during surgery¿.
 
Event Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CODMAN DISPOS PERFORATOR
Type of Device
DISPOSABLE PERFORATORS
Manufacturer (Section D)
RAYNHAM
325 paramount drive
325 paramount drive
raynham MA
Manufacturer (Section G)
RAYNHAM
325 paramount drive
raynham MA
Manufacturer Contact
vivian nelson
1100 campus drive
princeton, NJ 
6099362319
MDR Report Key16502930
MDR Text Key311042982
Report Number3014334038-2023-00023
Device Sequence Number1
Product Code HBF
UDI-Device Identifier10381780513599
UDI-Public10381780513599
Combination Product (y/n)N
PMA/PMN Number
K791101
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 05/18/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number261221
Device Catalogue Number261221
Device Lot Number6416521
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/13/2023
Initial Date FDA Received03/08/2023
Supplement Dates Manufacturer Received05/09/2023
Supplement Dates FDA Received05/18/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/22/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-