• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDOS INTERNATIONAL SARL CORACOID TOP HAT DRILL; CANNULATED SURGICAL DRILL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDOS INTERNATIONAL SARL CORACOID TOP HAT DRILL; CANNULATED SURGICAL DRILL Back to Search Results
Model Number 288202
Device Problem Degraded (1153)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/17/2023
Event Type  malfunction  
Event Description
This is report 1 of 3 for (b)(4).It was reported by a healthcare professional in australia that during an unknown arthroscopic procedure on (b)(6) 2023, it was observed that the coracoid top hat drill device was blunt.It was unknown how the procedure was completed.There were no adverse patient consequences nor surgical delay reported.No additional information was provided.
 
Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Additional narrative: udi: (b)(4).The device manufacture date is currently unavailable.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
 
Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: the actual device has been returned and is currently pending evaluation.Once reliability engineering evaluates the device, a supplemental medwatch report will be sent accordingly.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 correction narrative.D4: the lot number has been updated to reflect the correct information.Therefore, udi: (b)(4).
 
Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.The complaint device was received and evaluated.Upon visual inspection, it was found that the drill tip is dull.The device's shaft has no structural anomalies.Manufacturing record evaluation is not required as the reported event is not associated with the manufacturing process and/or the potential cause of the defect cannot be associated to manufacturing.As part of depuy synthes mitek quality process all devices are manufactured, inspected, and released to approved specifications.According with the visual inspection, this complaint can be confirmed.The possible root cause can be attributed to hard and repeated use, since this device is reusable, the continuous use and sterilization process can cause metal fatigue leading to a dull drill tip, however, this cannot be conclusively determined.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to monitor additional complaint information for potential safety signals through complaint trending as part of post market surveillance.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CORACOID TOP HAT DRILL
Type of Device
CANNULATED SURGICAL DRILL
Manufacturer (Section D)
MEDOS INTERNATIONAL SARL
chemin blanc 38
le locle CH-24 00
SZ  CH-2400
Manufacturer (Section G)
MEDOS INTERNATIONAL SARL
chemin-blanc 38
le locle CH-24 00
SZ   CH-2400
Manufacturer Contact
kate karberg
325 paramount drive
raynham, MA 02767
3035526892
MDR Report Key16504399
MDR Text Key311081007
Report Number1221934-2023-01085
Device Sequence Number1
Product Code HTW
UDI-Device Identifier10886705026807
UDI-Public10886705026807
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number288202
Device Catalogue Number288202
Device Lot Number20M04
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/17/2023
Initial Date FDA Received03/08/2023
Supplement Dates Manufacturer Received04/19/2023
05/14/2023
Supplement Dates FDA Received04/20/2023
05/15/2023
Was Device Evaluated by Manufacturer? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient SexMale
-
-