|
Model Number 71935069 |
Device Problem
Positioning Problem (3009)
|
Patient Problem
Rupture (2208)
|
Event Date 02/14/2023 |
Event Type
Injury
|
Event Description
|
It was reported that, during a meniscal root repair surgery, surgeon made four attempts to drill the tunnel using a cannulated drill pac and the acufex director mrr; however each time the dr missed the target spot by 5 mm.Surgery resumed after a delay greater than 30 minutes, by using a back-up cannulated drill which drilled exactly in the spot, however an additional bone hole was needed for these purpose, leaving a 2.8mm drill channels voids.Patient health status did not face any complication so far.No further complications were reported.
|
|
Manufacturer Narrative
|
Internal complaint reference: (b)(4).
|
|
Event Description
|
It was reported that, during a meniscal root repair surgery, surgeon made four attempts to drill the tunnel using a cannulated drill pac and the acufex director mrr; however each time the dr missed the target spot by 5 mm.Surgery resumed after a delay greater than 30 minutes, by using a back-up cannulated drill, which drilled exactly in the spot, however an additional bone hole was needed for these purpose, leaving three 2.8mm drill channels voids.Patient health status did not face any complication so far.No further complications were reported.
|
|
Manufacturer Narrative
|
B5.
|
|
Event Description
|
It was reported that, during a meniscal root repair surgery, surgeon made four attempts to drill the tunnel using a cannulated drill pac; however each time, the surgeon missed the target spot by 5 mm.Surgery resumed after a surgical delay greater than 30 minutes, by using a back-up cannulated drill which drilled exactly in the spot.No further complications were reported.
|
|
Manufacturer Narrative
|
B5 was updated.H3, h6: the reported device was received for evaluation.A visual inspection revealed the cutting surfaces of the drill bit were chipped and worn.One of the cannulated sleeves was deformed at the proximal end and spiral wear marks were noted on the exterior surface.A functional evaluation found the drill bit fit in the returned cannula sleeves and there was no warping or off-axis rotation noted.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A complaint history review found no similar reported events.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A clinical evaluation was performed and found no further risk to the patient is anticipated as a result of the additional bone holes.No patient complications have been noted following the procedure.Since no further harm is anticipated no further clinical assessment is warranted at this time.The complaint was confirmed and the root cause was associated with unintended use of the device.Factors that could have contributed to the reported event include application of inappropriate lateral force and/or use of a faulty concomitant device.No containment or corrective actions are recommended at this time.
|
|
Search Alerts/Recalls
|
|
|