Model Number MC2-5070R
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Metal Related Pathology (4530)
Event Date 03/02/2023
Event Type  Injury  
Manufacturer Narrative
The device has not been returned for evaluation.The root cause is unable to be determined at this time.A review of the device history records (dhr) confirmed the device met all quality inspections and specifications prior to release.If any additional information is provided, a supplemental report will be submitted.
 
Event Description
Information was received that upon removal there was metallosis present.
 
Event Description
Information was received that during removal for definitive fusion after treatment goals were achieved, it was noted there was "metallosis" present.
 
Manufacturer Narrative
The device has not been returned to nuvasive for evaluation.Investigation has been conducted using provided x-ray images.It was noted that the rod had been partially distracted.There were no obvious device-related abnormalities noted.Based on the provided information, no objective evidence has been found to indicate the presence of metallosis.The presence of metallosis cannot be verified using the radiographic pictures or x-ray images that have been provided.The reported event cannot be confirmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAGEC GENERATION 2B SPINAL BRACING AND DISTRACTION SYSTEM
Type of Device
GROWING ROD SYSTEM - MAGNETIC ACTUATION
Manufacturer (Section D)
NUVASIVE SPECIALIZED ORTHOPEDICS, INC.
101 enterprise dr. ste 100
aliso viejo CA 92656
Manufacturer (Section G)
NUVASIVE SPECIALIZED ORTHOPEDICS, INC.
101 enterprise dr. ste 100
aliso viejo CA 92656
Manufacturer Contact
geoff gannon
101 enterprise dr. ste 100
aliso viejo, CA 92656
8583448112
MDR Report Key16513658
MDR Text Key310982342
Report Number3006179046-2023-00155
Device Sequence Number1
Product Code PGN
UDI-Device Identifier00887517834010
UDI-Public887517834010
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K171791
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/31/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberMC2-5070R
Device Lot Number7090903AAA
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 03/02/2023
Initial Date FDA Received03/09/2023
Supplement Dates Manufacturer Received03/30/2023
Supplement Dates FDA Received03/31/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/09/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention; Other;
Patient Age11 YR
Patient SexFemale
Patient Weight43 KG
-
-