VYAIRE MEDICAL INC. 3100 HIGH FREQUENCY OSCILLATING VENTILATOR (HFOV); VENTILATOR, HIGH FREQUENCY Back to Search Results
Model Number 3100A VENT,115V,60HZ,DOM
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/14/2023
Event Type  Injury  
Manufacturer Narrative
A vyaire field service engineer checked the unit and ran performance check.Found delta p at 82.Calibrated ddi (dynamic displacement indicator) and pr2 (pressure regulator).Delta p within specifications at 67 afterwards.Vyaire medical proceeded with completing 2k pm.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
 
Event Description
It was reported to vyaire medical problem with 3100a ventilator where the customer observed that the patient was not improving, so they swapped the unit out with another 3100a ventilator.They just want to make sure that the unit is functioning properly.Furthermore, there was no patient harm reported associated with the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3100 HIGH FREQUENCY OSCILLATING VENTILATOR (HFOV)
Type of Device
VENTILATOR, HIGH FREQUENCY
Manufacturer (Section D)
VYAIRE MEDICAL INC.
26125 n. riverwoods blvd.
mettawa IL 60045
Manufacturer (Section G)
VYAIRE MEDICAL INC.
1100 bird center drive
palm springs CA 92262
Manufacturer Contact
erika bonilla
510 technology drive
irvine, CA 92618
MDR Report Key16513760
MDR Text Key311043483
Report Number2021710-2023-17391
Device Sequence Number1
Product Code LSZ
UDI-Device Identifier10846446003079
UDI-Public(01)10846446003079(11)20071113
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P890057
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 03/09/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number3100A VENT,115V,60HZ,DOM
Device Catalogue Number768901
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/14/2023
Initial Date FDA Received03/09/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/13/2007
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-