Model Number 9-AVP2-022
Device Problems Difficult to Fold, Unfold or Collapse (1254); Improper or Incorrect Procedure or Method (2017)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/10/2023
Event Type  malfunction  
Event Description
It was reported that on (b)(6) 2023, an 22mm amplatzer vascular plug ii was chosen for implant with a non-abbott delivery system.It was reported the plug exhibited an "elongating" abnormal shape during deployment.The device was recaptured and repositioned but maintained its elongation shape.A decision was made to continue with the procedure and the device was implanted in the patient.The physician could not confirm if the elongation shape was due to patient anatomy.The patient remained hemodynamically stable throughout the procedure and their status was reported as stable.There was no report of any interaction with cardiac structures during deployment or any angulation/kink noticed with the delivery system.
 
Manufacturer Narrative
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Manufacturer Narrative
An event of device deformity during deployment was reported.A returned device assessment could not be performed as the device was not returned for analysis.Possible causes of device deformity include use of the incorrect size delivery system, anatomical interference, and angulation or kink in the delivery system upon deployment.Information from the field indicated that there were no kinks noticed in the delivery system and there was no interaction with cardiac structures during the deployment.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all defined manufacturing specifications.Based on the information received, the cause of the reported incident could not be conclusively determined.The field indicated that the device was implanted in it's elongated state.Please note per the instructions for use, if device position is unsatisfactory: stabilize the wire and re-advance the delivery catheter until the device is completely within the catheter.Reposition and deploy, or remove the device from the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMPLATZER VASCULAR PLUG II
Type of Device
DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ABBOTT MEDICAL REG# 2135147
5050 nathan ln n
plymouth MN 55442
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key16516615
MDR Text Key311041096
Report Number2135147-2023-01092
Device Sequence Number1
Product Code KRD
UDI-Device Identifier00811806010441
UDI-Public00811806010441
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K071699
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 05/17/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number9-AVP2-022
Device Catalogue Number9-AVP2-022
Device Lot Number7935892
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/15/2023
Initial Date FDA Received03/09/2023
Supplement Dates Manufacturer Received05/02/2023
Supplement Dates FDA Received05/17/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/20/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-