THE SPECTRANETICS CORPORATION SPECTRANETICS LEAD LOCKING DEVICE; STYLET, CATHETER Back to Search Results
Model Number 518-062
Device Problem Material Separation (1562)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/21/2023
Event Type  Injury  
Manufacturer Narrative
Patient's date of birth unk.Patient''s weight unk.Other relevant history unk.Device lot number, expiration date unk.The device was discarded, thus no investigation could be completed.Device manufacture date unk because lot number unk.Although device damage is a known risk of complication with use of the lld, the cause for the lld breaking could not be established.Submission of this report does not, in itself, represent a conclusion by the manufacturer and/or authorized representative or the national competent authority that the content of this report is complete or accurate, that the medical device(s) listed failed in any manner and/or that the medical device(s) caused or contributed to an alleged death or deterioration in the state of the health of any person.
 
Event Description
A lead extraction procedure commenced to remove a right ventricular (rv) and a right atrial (ra) lead due to malfunction.Spectranetics lead locking devices (llds) were inserted into each lead, along with suture, to provide traction.Using a spectranetics 11f tightrail guardian motorized dilator sheath, the rv lead was removed successfully.While working to remove the ra lead with the tightrail guardian and traction, the lead came back into the subclavian where it got stuck.While attempting to remove the lead with the sheath and while significant traction was being applied, the ra lead/lld separated.The ra lead/lld remnants were removed via femoral snare, and the procedure was completed with no reported patient harm.This event captures the lld which separated along with the ra lead, requiring intervention.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPECTRANETICS LEAD LOCKING DEVICE
Type of Device
STYLET, CATHETER
Manufacturer (Section D)
THE SPECTRANETICS CORPORATION
9965 federal drive
colorado springs CO 80921
Manufacturer (Section G)
SPECTRANETICS
9965 federal drive
colorado springs CO 80921
Manufacturer Contact
barbara creel
9965 federal drive
colorado springs, CO 80921
719447-246
MDR Report Key16518043
MDR Text Key311046157
Report Number1721279-2023-00046
Device Sequence Number1
Product Code DRB
UDI-Device Identifier00813132023072
UDI-Public00813132023072
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K142116
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 03/09/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number518-062
Device Catalogue Number518-062
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received03/09/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
BOSTON SCIENTIFIC 4470 RA PACING LEAD; BOSTON SCIENTIFIC 5076 RV PACING LEAD; SPECTRANETICS LEAD LOCKING DEVICE IN RV LEAD; SPECTRANETICS TIGHTRAIL GUARDIAN DILATOR SHEATH
Patient Outcome(s) Required Intervention;
Patient Age77 YR
Patient SexMale
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-