E2, e3: multiple attempts were made to obtain additional information but were unsuccessful.This report is being supplemented to provide additional information based on the legal manufacturer's device evaluation and investigation.Based on the results of the device evaluation, several non-reportable phenomena such as the fan making a grinding noise, a damaged front panel, a bent front panel chassis, a broken dy switch, a scope socket locking mechanism that doesn¿t protect the terminals, corrosion inside the air pump tube, and a unit with a missing lamp were confirmed.The reported event (the fan not working) was confirmed.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 8 years since the subject device was manufactured.Based on the results of the legal manufacturer's investigation, a definitive root cause of the fan not working could not be identified.However, it¿s likely the cause is related to a malfunction of the fan.Olympus will continue to monitor field performance for this device.
|