Model Number G5S-11A |
Device Problem
Incorrect Interpretation of Signal (1543)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/26/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
Zoll medical corporation has not received the product for evaluation and this complaint is still under investigation.
|
|
Event Description
|
Complainant alleged that while attempting to treat a patient (age & gender unknown), the device gave a "no shock advised" prompt for a rhythm clinicians believed was shockable.Complainant indicated that the clinician obtained another device to continue treating the patient.Complainant indicated that there was no adverse effect to the patient due to the reported malfunction.
|
|
Manufacturer Narrative
|
The device was not returned to zoll medical corporation.Instead, the clinical file of the customer's report was provided.Review of the clinical file from the reported event concluded that the device worked as designed and within the limitations of the technology.A review of the unit's configuration concluded the unit operated as intended.Analysis of reports of this type has not identified an increase in trend.
|
|