Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Additional narrative: 510k: this report is for an unk - constructs: expedium/unknown lot.Part and lot numbers are unknown; udi number is unknown.Complainant part is not expected to be returned for manufacturer review/investigation.Reporter is a j&j employee.Investigation summary: product was not returned.Based on the information available, it has been determined that no corrective and preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
This report is being filed after the review of a clinical evaluation report (cer) from a database related research activity (drra) between january 1, 2017 to march 31, 2022 with a total of 141 patients using confidence¿ spinal cement with the viper¿ or expedium¿ fenestrated screw systems¿.The following complications were reported as follows: (n=9) reoperation within 365 days post-index surgery.(n=8) revision within 365 days post-index surgery.This report is for an unknown depuyspine confidence¿ spinal cement with the viper¿ or expedium¿ fenestrated screw systems¿.This report is for one unk - constructs: expedium.This is report 2 of 3 for complaint (b)(4).
|