The reported event was unconfirmed as the problem could not be reproduced.The ureteral stent was returned in the opened original packaging.The only curves that could be seen are the two pigtails that are formed and the expected curve from the way the stents are packaged and shipped.No defects could be confirmed.The sample passed all dimensional requirements; the outer diameter measured 0.0612" with a laser micrometer, the tip inner diameter measured 0.039" with a pin gauge, the overall length measured 259mm, the proximal pigtail outer diameter measured 0.5760", and the distal pigtail outer diameter measured 0.5845".A 0.035" guidewire successfully passed through the sample with no resistance.As no patient involvement was reported the device was not used, however, is intended for treatment purposes.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.As the reported event is unconfirmed, a label/packaging review is not required.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text: the actual/suspected device was inspected.
|