The initial reporter stated they received discrepant results for one patient sample tested with ldhi2 lactate dehydrogenase acc.To ifcc ver.2 on a cobas integra 400 plus (serial number (b)(4).No questionable results were reported outside of the laboratory.The customer initially had issues with an unspecified number of patient samples tested with ldh.The results would be high and then repeat lower.The probes were replaced and precision was checked.A wash station was also inspected.The field application specialist updated the application settings to use a different application of the same reagent, but the issue still persisted leading to the complained sample results.The sample initially resulted in an ldh value of 733 u/l.Due to ongoing issues of the same nature, the sample was repeated twice, resulting in values of 211 u/l and 217 u/l.
|