MAUDE Adverse Event Report: BIOTRONIK SE & CO. KG ENITRA 8 SR-T; PACEMAKER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOTRONIK SE & CO. KG ENITRA 8 SR-T; PACEMAKER Back to Search Results
Model Number 407159
Device Problems Failure to Capture (1081); High impedance (1291); High Capture Threshold (3266)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/06/2023
Event Type  malfunction  
Event Description
The device was explanted due to loss of capture, high pacing threshold and high impedance.No adverse patient events were reported.Should additional information be received, this file will be updated.
 
Manufacturer Narrative
The pacemaker was received for analysis.Prior to the analysis of the device, the quality documents accompanying the manufacturing process for this pacemaker were re-investigated and all production steps were performed accordingly.Particularly the final acceptance test proved the device functions to be as specified.Upon receipt, the pacemaker was interrogated, and the memory content was analyzed indicating no pacemaker malfunction.Therefore, the header of the device was analyzed.The set screw could be easily screwed in and out, there was no foreign material inside the header bore.All dimensions of the header bore were within the range requested by the standard specifications.Also, the spring element of the pacemaker did not show any deviations.The ability of the device to deliver therapies was verified.The anti-bradycardia pacing pulses proved to be normal and in amplitude and frequency as programmed.There was no indication of a device malfunction.An additional long-term pacing test was initiated.During the test, each pacing pulse was recorded.The evaluation of these pacing pulses documented regular device behavior.There was no intermittent or permanent loss of output.In conclusion, the device is fully functional.The analysis did not show any deviations from the technical specifications.The analysis did not reveal any sign of a material or manufacturing problem.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENITRA 8 SR-T
Type of Device
PACEMAKER
Manufacturer (Section D)
BIOTRONIK SE & CO. KG
woermannkehre 1
berlin 12359
Manufacturer Contact
6024 jean road
lake oswego, OR 97035
8772459800
MDR Report Key16534154
MDR Text Key311237907
Report Number1028232-2023-01238
Device Sequence Number1
Product Code NVZ
Combination Product (y/n)N
PMA/PMN Number
P950037
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 06/23/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/31/2024
Device Model Number407159
Device Catalogue NumberSEE MODEL NO.
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/28/2023
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/10/2023
Initial Date FDA Received03/13/2023
Supplement Dates Manufacturer Received06/22/2023
Supplement Dates FDA Received06/23/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/04/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization;
-
-