Model Number UHI-4
Device Problem Failure to Clean Adequately (4048)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/11/2023
Event Type  malfunction  
Event Description
The customer reported that, during the routine reprocessing of the customer¿s high flow insufflation unit, it was reported that the device had no air flow.Upon inspection and testing of the returned unit, it was discovered that the carbon dioxide gas channel was completely obstructed by foreign matter, which prevented insufflation.There were no reports of patient or user harm associated with this event.This medical device report (mdr) is being submitted to capture the reportable malfunction found during evaluation.
 
Manufacturer Narrative
The device was returned to an olympus repair facility, and an evaluation of the device was performed.During the evaluation, it was discovered that the carbon dioxide gas channel was completely obstructed by foreign matter.The customer's originally reported issue was confirmed.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the carbon dioxide (co2) gas pipeline was caused by blocked by dirt, resulting in no gas insufflation.The cause of the foreign matter could not be identified.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIGH FLOW INSUFFLATION UNIT
Type of Device
HIGH FLOW INSUFFLATION UNIT
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key16534705
MDR Text Key311579631
Report Number3002808148-2023-02443
Device Sequence Number1
Product Code HIF
UDI-Device Identifier04953170324147
UDI-Public04953170324147
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K122180
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 04/14/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberUHI-4
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/14/2023
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 02/11/2023
Initial Date FDA Received03/13/2023
Supplement Dates Manufacturer Received03/23/2023
Supplement Dates FDA Received04/14/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/10/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
OTV-S190
-
-