• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC CRYOCATH LP ARCTIC FRONT ADVANCE PRO¿ CARDIAC CRYOABLATION CATHETER; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC CRYOCATH LP ARCTIC FRONT ADVANCE PRO¿ CARDIAC CRYOABLATION CATHETER; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION Back to Search Results
Model Number AFAPRO28
Device Problems Mechanical Problem (1384); Infusion or Flow Problem (2964); Material Deformation (2976); Material Twisted/Bent (2981)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/08/2022
Event Type  malfunction  
Manufacturer Narrative
Product event summary: the afapro28 balloon catheter with lot number 16380 was returned and analyzed.External visual inspection of the balloon segment was performed no anomalies were identified.The balloons and sleeve are all intact with no apparent issues.The catheter smart chip data was downloaded and reviewed.Data indicated the catheter was used for eight applications on the reported event date.The catheter was recognized and passed the electrical integrity verifications and performance test.The catheter completed the inflation, ablation, and thawing phases with no console system notices generated.No performance issues were identified.All pressure and flow were in range and temperature curve had no oscillation or overshoot.A guide wire lumen kink was observed inside the balloon during an inflation.During inspection of the shaft segment, a guide wire lumen kink/twist was observed 1.17 inches proximal to the catheter tip.Pressure testing did not identify any leakage from the kink.In conclusion, the balloon catheter failed the returned product inspection due to the guide wire lumen kink.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that during a cryo ablation procedure, the balloon catheter inflation was not as expected.On fluoroscopy, it was noted that the shaft of the inner portion of the balloon was bent.The balloon catheter was removed and examined; the balloon inner shaft was kinked and had a "strange shape." the balloon catheter was replaced which resolved the issue. the case was completed with cryo. no patient complications have been reported as a result of this event.On (b)(6) 2023 after a completed case, the balloon catheter subsequently tested out of specification per the manufacturer's investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARCTIC FRONT ADVANCE PRO¿ CARDIAC CRYOABLATION CATHETER
Type of Device
CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION
Manufacturer (Section D)
MEDTRONIC CRYOCATH LP
9000 autoroute transcanadienne
pointe-claire,qc H9R 5 Z8
CA  H9R 5Z8
Manufacturer (Section G)
MEDTRONIC CRYOCATH LP
9000 autoroute transcanadienne
pointe-claire,qc H9R 5 Z8
CA   H9R 5Z8
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key16537657
MDR Text Key311245402
Report Number3002648230-2023-00095
Device Sequence Number1
Product Code OAE
UDI-Device Identifier00763000431464
UDI-Public00763000431464
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P100010/S015
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 03/14/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/26/2023
Device Model NumberAFAPRO28
Device Catalogue NumberAFAPRO28
Device Lot Number16380
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/13/2022
Initial Date Manufacturer Received 02/14/2023
Initial Date FDA Received03/14/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/26/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-