Model Number BI70002000
Device Problems Adverse Event Without Identified Device or Use Problem (2993); Unexpected Shutdown (4019); Application Program Freezes, Becomes Nonfunctional (4031)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/20/2023
Event Type  malfunction  
Event Description
Medtronic received information regarding an imaging system.It was reported that the system was unexpectedly shutting down when 3d scan was initiated.Ap and lateral images were taken and the image acquisition system (ias) pendant went black and shut down when attempting to take 3d scan.The system was plugged into its own outlet and was fully charged.There was novisible damage to the mobile viewing station  (mvs) power cord or umbilical cable.The handswitch was being used and it was noted that 3d scan was eventually possible.There was no impact on patient outcome and unknown surgical delay.
 
Manufacturer Narrative
Concomitant medical products: other relevant device(s) are: product id: bi71000230; product id: bi71000576; product id: bi71000424.A medtronic representative went to the site to test the equipment.Testing found that there were errors with the starter or booster.The starter board and booster board were replaced.The broken dongle and bent key was corrected.A system checkout was performed and the system is working as intended.No parts have been returned to the manufacturer for analysis.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
H3: hardware analysis was completed on the returned rear cab assembly and it passed continuity testing and visual inspection.No key or dongle was returned with the assembly.H6: b01, c19 and d14 apply to the returned concomitant product id: kit svc o2 bi71000424 panel rearcab brkt, lot number 52576 rev.E.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
H3: the bi71000230r (ln: s1829gos rev.B) was returned for analysis.Analysis found no fault with the returned generator.The generator was installed into a known good system and the system functioned as expected.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
H3: the starter board was returned to the manufacturer for analysis.Analysis found that with the starter installed into a test system, the system began beeping during boot up and failed to fully boot.The starter board was determined to be faulty.Analysis found that the reported event was related to an electrical issue.H6: fdm b01, fdr c02, and fdc d02 are applicable to the hardware analysis.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
O-ARM O2 IMAGING SYSTEM
Type of Device
INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM
Manufacturer (Section D)
SANMINA -SCI SYSTEMS
km 15.5 no. 29, plant 06
carr. chapala-guadalajara,jal 45640
MX  45640
Manufacturer (Section G)
SANMINA -SCI SYSTEMS
km 15.5 no. 29, plant 06
carr. chapala-guadalajara,jal 45640
MX   45640
Manufacturer Contact
glen belmer
7000 central avenue ne rcw215
minneapolis, MN 55432
6122713209
MDR Report Key16542662
MDR Text Key311676430
Report Number3006544299-2023-00133
Device Sequence Number1
Product Code OWB
UDI-Device Identifier00763000132378
UDI-Public00763000132378
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K200074
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup
Report Date 05/25/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberBI70002000
Device Catalogue NumberBI70002000
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/20/2023
Initial Date FDA Received03/15/2023
Supplement Dates Manufacturer Received04/06/2023
04/28/2023
05/20/2023
Supplement Dates FDA Received04/28/2023
05/16/2023
05/26/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/22/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-