The product investigation was completed.Device evaluation details: visual analysis revealed reddish material inside the pebax.A screening test was performed, and the device was visualized and recognized correctly; however, the force vector was observed inverted and hi force was displayed.A sem (scanning electron microscope) study was performed, and mechanical damage and a hole was observed on the pebax surface.Since the sensor values were found within specifications, this issue could be related to the reddish material inside the pebax, however, the root cause of the damage on the pebax could not be determined, but it was concluded that it occurred outside the bwi manufacturing facilities.A manufacturing record evaluation was performed for the finished device 30868546m, and no internal action was found during the review.The issue reported by the customer was confirmed.It should be noted that product failure is multifactorial.The instructions for use contain the following recommendations: the force sensor of the catheter is disconnected.If the problem persists, replace the catheter cable or the catheter.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's reference number: (b)(4).
|
A patient underwent an ablation procedure with a thermocool® smart touch¿ electrophysiology catheter for which biosense webster¿s product analysis lab revealed reddish material inside the pebax.A screening test was performed, and the device was visualized and recognized correctly; however, the force vector was observed inverted and hi force was displayed.A sem (scanning electron microscope) study was performed, and mechanical damage and a hole was observed on the pebax surface.During the procedure, the force issue was originally identified.It was reported that during procedure, the force value could not be zeroed.A second device was used to complete the operation.There was no adverse event reported on patient.Force issue is not mdr-reportable.Hole in the pebax is mdr-reportable.
|