Model Number KNEE SCORPION |
Device Problem
Break (1069)
|
Patient Problems
Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/20/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
|
|
Event Description
|
On (b)(6) 2023, it was reported by a sales representative via sems that an ar-12990 knee scorpions' top arm broke off.This occurred during a meniscal root repair, and all fragments of the broken device were retrieved.The case was completed using only one suture versus two.There was no additional information provided.Additional information has been requested.
|
|
Manufacturer Narrative
|
Complaint confirmed.Upon visual inspection, it was noted that the top (moving) jaw of the device was broken off.The broken piece was not returned for investigation.The cause of this remains undetermined, however, probable cause can be attributed to prying/leveraging forces being used on the device.
|
|