• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. KNEE SCORPION; MANUAL INSTR, GENERAL SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. KNEE SCORPION; MANUAL INSTR, GENERAL SURGICAL Back to Search Results
Model Number KNEE SCORPION
Device Problem Break (1069)
Patient Problems Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/20/2023
Event Type  malfunction  
Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
 
Event Description
On (b)(6) 2023, it was reported by a sales representative via sems that an ar-12990 knee scorpions' top arm broke off.This occurred during a meniscal root repair, and all fragments of the broken device were retrieved.The case was completed using only one suture versus two.There was no additional information provided.Additional information has been requested.
 
Manufacturer Narrative
Complaint confirmed.Upon visual inspection, it was noted that the top (moving) jaw of the device was broken off.The broken piece was not returned for investigation.The cause of this remains undetermined, however, probable cause can be attributed to prying/leveraging forces being used on the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KNEE SCORPION
Type of Device
MANUAL INSTR, GENERAL SURGICAL
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
vik bajnath
8009337001
MDR Report Key16544756
MDR Text Key311416261
Report Number1220246-2023-06419
Device Sequence Number1
Product Code NBH
UDI-Device Identifier00888867196322
UDI-Public00888867196322
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Distributor
Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date 05/08/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberKNEE SCORPION
Device Catalogue NumberAR-12990
Device Lot Number76416
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 02/20/2023
Initial Date FDA Received03/15/2023
Supplement Dates Manufacturer Received02/20/2023
Supplement Dates FDA Received05/08/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/08/2016
Is the Device Single Use? No
Type of Device Usage Initial
Patient Sequence Number1
-
-