Catalog Number A100400000000
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/09/2023
Event Type  malfunction  
Manufacturer Narrative
There has been a previous report received with a similar device where this malfunction resulted in a serious injury.Therefore, it must be presumed that recurrence of this malfunction could possibly cause or contribute to a serious injury or require medical or surgical intervention to preclude such.As such, this event is reportable per 21cfr part 803.Investigation results received.As per service engineer- h1004c5210150 - cartridge x s-problem.H1004c5210500 - head cap -problem.H1004c5470220 - neck-x- problem.Under warranty invoice no.Sr/21-22/75 dated 14-07-2021.Investigation result -cartridge neck and cap defective - replaced.Any additional information received will be submitted as it becomes available.
 
Event Description
This event it is reported that x-smart w/contra angle eur won't hold files.No injury occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
X-SMART W/CONTRA ANGLE EUR
Type of Device
HANDPIECE, DIRECT DRIVE, AC-POWERED
Manufacturer (Section D)
MAILLEFER INSTRUMENTS HOLDING SARL
ch. du verger 3
ballaigues vaud CH-13 38
SZ  CH-1338
Manufacturer (Section G)
MAILLEFER INSTRUMENTS HOLDING SARL
ch. du verger 3
ballaigues vaud CH-13 38
SZ   CH-1338
Manufacturer Contact
hannah seevaratnam
221 west philadelphia st.
york, PA 17401
7178457511
MDR Report Key16546594
MDR Text Key311483702
Report Number8031010-2023-00765
Device Sequence Number1
Product Code EKX
Combination Product (y/n)N
Reporter Country CodeIN
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Dentist
Type of Report Initial
Report Date 03/15/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberA100400000000
Device Lot Number0A744819
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Distributor Facility Aware Date03/10/2023
Initial Date Manufacturer Received 03/10/2023
Initial Date FDA Received03/15/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Type of Device Usage A
Patient Sequence Number1
-
-