SHIRAKAWA OLYMPUS CO., LTD. CYSTO-NEPHRO VIDEOSCOPE Back to Search Results
Model Number CYF-VH
Device Problem Material Integrity Problem (2978)
Patient Problem Foreign Body In Patient (2687)
Event Date 02/13/2023
Event Type  Injury  
Event Description
The customer reported to olympus that black dots or fragments went into the patient's bladder while using the cysto-nephro videoscope during an unknown procedure.Additional information has been requested but no further details have been provided at this time.
 
Manufacturer Narrative
The suspect device has been returned to olympus for evaluation.The user reported problem was not confirmed.However, the bending section cover adhesive was noted cracked.Replacement of the bending rubber was recommended.Additional information is being requested.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation or if additional information is received.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 3 years since the subject device was manufactured.Based on the results of the investigation, the reported issue could not be reproduced.Therefore, a relationship between the subject device and adverse event could not be determined.The customer did not provide the requested details regarding their reprocessing steps.The event can be detected/prevented by following the instructions for use (ifu) in sections: ¿ifu: operation manual chapter 5 troubleshooting, 5.1 troubleshooting.¿ this supplemental report includes information added to d8.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CYSTO-NEPHRO VIDEOSCOPE
Type of Device
CYSTO-NEPHRO VIDEOSCOPE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key16547973
MDR Text Key311343767
Report Number3002808148-2023-02531
Device Sequence Number1
Product Code FAJ
UDI-Device Identifier04953170310461
UDI-Public04953170310461
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
K221683
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 06/19/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCYF-VH
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/15/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 02/13/2023
Initial Date FDA Received03/15/2023
Supplement Dates Manufacturer Received06/14/2023
Supplement Dates FDA Received06/19/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/08/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-