Model Number 3000TFX |
Device Problems
Calcified (1077); Gradient Increase (1270); Perivalvular Leak (1457); Insufficient Information (3190)
|
Patient Problems
Angina (1710); Insufficient Information (4580)
|
Event Date 01/18/2023 |
Event Type
Injury
|
Event Description
|
It was learned via the patient registry that a 23mm 3000tfx pericardial aortic valve was disabled via valve-in-valve procedure after an implant duration of nine (9) years, five (5) months due to unknown reasons.The tavr procedure was performed with a 23mm 9755rsl transcatheter valve.Edwards lifesciences maintains an implant patient registry.This registry is a patient tracking mechanism for serialized edwards implantable devices (bioprosthetic heart valves and annuloplasty rings), rather than a true post-market surveillance registry.Through the registry, edwards is notified when these devices are implanted.In addition, patient and/or device status may be reported to the registry via the implantation data cards.The information is received from various sources (e.G.Surgeon, hospital, and patient family members) and is not received in the form of a conventional customer complaint.The information reported may or may not be related to the edwards device.
|
|
Manufacturer Narrative
|
The subject device is not available for evaluation, as it remains implanted in the patient.The device history record was reviewed and shows that this device met all manufacturing specifications for product release prior to distribution.No issues were identified that would have impacted this event.The investigation is still in progress; therefore, a conclusion has yet to be established.A supplemental report will be submitted accordingly upon investigation completion.Edwards will continue to review and monitor all reported events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
|
|
Manufacturer Narrative
|
Calcific degeneration is a common cause of bioprosthetic heart valve failures.Many factors contribute to the onset and propagation of calcification.These include patient factors (age, disease state, pharmacological intervention, etc.), mechanical stress related to the valve's hemodynamic performance, and glutaraldehyde fixation of tissue.Of these, the fixation process is a relatively minor contributor to calcification for edwards' tissue valves due to anti-calcification treatments during manufacturing.Though numerous studies have been conducted on preventive calcification strategies in bioprosthetic heart valves, the causes of calcification are not fully understood and there are still no mechanisms or medical therapies which fully prevent bioprostheses from calcifying.Through further investigation, it was determined that the root cause of this event was most likely due to patient related factors.
|
|
Event Description
|
It was learned via the patient registry that a 23mm 3000tfx pericardial aortic valve was disabled via valve-in-valve procedure after an implant duration of nine (9) years, five (5) months due to severe stenosis, moderate ar likely due to calcification (leaflets not well visualized on echo).The patient presented with progressive angina.The tavr procedure was performed with a 23mm 9755rsl transcatheter valve.
|
|