HEARTSINE TECHNOLOGIES LTD PACKAGE, 350P,PP03J,JA,350-STR-JA-10; AUTOMATED EXTERNAL DEFIBRILLATOR Back to Search Results
Model Number 350P
Device Problem Electrical /Electronic Property Problem (1198)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/21/2022
Event Type  malfunction  
Manufacturer Narrative
The customer contacted heartsine to report a non-critical issue with their device.During evaluation, a pogo pin failure was found.This had resulted in the voltage fluctuations and low battery fails.The device was scrapped by heartsine and the customer was provided with a replacement device.
 
Event Description
The customer contacted heartsine to report a non-critical issue with their device.During evaluation, a pogo pin failure was found.This may prevent the device from powering on which may prevent or delay defibrillation therapy.There was no report of patient use associated to the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PACKAGE, 350P,PP03J,JA,350-STR-JA-10
Type of Device
AUTOMATED EXTERNAL DEFIBRILLATOR
Manufacturer (Section D)
HEARTSINE TECHNOLOGIES LTD
207 airport road west
belfast BT3 9 ED
EI  BT3 9ED
Manufacturer (Section G)
HEARTSINE TECHNOLOGIES LTD
207 airport road west
belfast BT3 9 ED
EI   BT3 9ED
Manufacturer Contact
todd bandy
207 airport road west
belfast BT3 9-ED
EI   BT3 9ED
4258674577
MDR Report Key16551391
MDR Text Key311381940
Report Number3004123209-2023-00026
Device Sequence Number1
Product Code NSA
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
P160008
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 03/16/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number350P
Device Catalogue Number350-STR-JA-10
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/23/2023
Initial Date Manufacturer Received 02/24/2023
Initial Date FDA Received03/16/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/05/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-