HEARTSINE TECHNOLOGIES LTD PACKAGE, 350P,PP03J,JA,350-STR-JA-10; AUTOMATED EXTERNAL DEFIBRILLATOR
|
Back to Search Results |
|
Model Number 350P |
Device Problem
Electrical /Electronic Property Problem (1198)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/21/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
The customer contacted heartsine to report a non-critical issue with their device.During evaluation, a pogo pin failure was found.This had resulted in the voltage fluctuations and low battery fails.The device was scrapped by heartsine and the customer was provided with a replacement device.
|
|
Event Description
|
The customer contacted heartsine to report a non-critical issue with their device.During evaluation, a pogo pin failure was found.This may prevent the device from powering on which may prevent or delay defibrillation therapy.There was no report of patient use associated to the reported event.
|
|