LIVANOVA USA, INC. LEAD MODEL 302 Back to Search Results
Model Number 302-20
Device Problems Fracture (1260); High impedance (1291)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/20/2023
Event Type  malfunction  
Manufacturer Narrative
Livanova usa, inc.Submits this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation, based on information that livanova has obtained, but may not have been able to investigate or verify prior to the date the report was required by the fda.This report does not constitute an admission, or a conclusion by fda or anyone else, that the device, livanova, or livanova's employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects¿ or "malfunctions¿.These words are incorporated into the fda 3500a medwatch form by the fda, and livanova objects to their use.
 
Event Description
It was reported that patient was seen with high impedance.Further information was received from the nurse practitioner that x rays have not been taken and patient will be referred to surgeon.No known relevant surgical intervention has occurred to date.No other relevant information has been received to date.
 
Manufacturer Narrative
B3.Corrected data, event date; initial report inadvertently listed wrong date.
 
Event Description
The event date is 02/20/2023, the initial mdr submitted the wrong event date in error.
 
Event Description
It was later reported patient had lead revision surgery.During surgery, the surgeon noticed that the electrodes were not placed on the vagus nerve but were wrapped around something else, possibly the ansa cervicalis nerve.The surgeon also noted a clear break in the lead insulation in the upper portion of the lead, which would have been the cause of the high impedance.Following the surgery it was noted that the explanted lead serial number did not match the device on record.The explanted device has been received into product analysis.No other relevant information has been received to date.
 
Event Description
Product analysis has been completed and the fracture was confirmed.No other relevant information has been received to date.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEAD MODEL 302
Type of Device
LEAD
Manufacturer (Section D)
LIVANOVA USA, INC.
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
LIVANOVA USA, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
cindy scott
100 cyberonics blvd
suite 600
houston, TX 77058
2816672681
MDR Report Key16553634
MDR Text Key311675254
Report Number1644487-2023-00309
Device Sequence Number1
Product Code MUZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Nurse Practitioner
Type of Report Initial,Followup,Followup,Followup
Report Date 12/19/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date07/20/2009
Device Model Number302-20
Device Lot Number1538
Was Device Available for Evaluation? No
Event Location Other
Initial Date Manufacturer Received 03/16/2023
Initial Date FDA Received03/16/2023
Supplement Dates Manufacturer Received03/16/2023
10/09/2023
12/08/2023
Supplement Dates FDA Received03/16/2023
11/03/2023
12/19/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/01/2006
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age67 YR
Patient SexFemale
-
-