Model Number AB2000
Device Problems Visual Prompts will not Clear (2281); Connection Problem (2900); Incomplete or Inadequate Connection (4037)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/16/2023
Event Type  malfunction  
Event Description
A male patient underwent an aquablation procedure for symptomatic benign prostatic hyperplasia (bph).Procept biorobotics corporation (procept) became aware that during the aquablation procedure, upon setting up the aquabeam handpiece, the conformal planning unit (cpu) displayed a fault indicating a failed connection between the handpiece and the aquabeam motorpack.During troubleshooting, an additional handpiece was used in an effort to clear the fault, but the issue persisted, and the handpiece and motorpack were not engaging despite multiple troubleshooting steps.As a result, the aquablation procedure was aborted.There were no adverse health consequences to the patient due to this event.
 
Manufacturer Narrative
Adverse event problem component code 4756: per the instructions for use, the aquabeam motorpack, a re-usable component of the aquabeam robotic system, provides power to the aquabeam handpiece by means of dc motors.Submission of this report does not constitute an admission that the manufacturer's product caused or contributed to the event.
 
Manufacturer Narrative
The motorpack was returned for investigation.Visual inspection of the returned motorpack did not reveal any anomalies.Functional testing of the motorpack confirmed the reported failure as the connection between the handpiece and the motorpack was observed to be unsuccessful.The issue was isolated to a defective/damaged cable.Root cause is undeterminable as it is not clear how the cable was damaged.A review of the device history record (dhr) ab2000-b/ serial number (b)(6), and aquabeam motorpack with lot number 22c01828 was conducted, which confirmed that there were no non-conformances, failures, discrepancies, or missed steps during the manufacturing process that could be related to the reported event.The review indicated that the system met all design and manufacturing specifications when released for distribution.The current user manual um0101-00 rev.F, aquabeam robotic system user manual, us, english was reviewed.Dock the motorpack to the aquabeam handpiece: -verify aquabeam handpiece nozzle position (i.E.The movement of the blue led) homes to the base of the aquabeamhandpiece (fully proximal).-apply sterile tape over the connection between aquabeam handpiece and motorpack to seal it.-keep the motorpack and the aquabeam handpiece assembly with the scope clamp assembly in a secure and sterile environment.Submission of this report does not constitute an admission that the manufacturer's product caused or contributed to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AQUABEAM ROBOTIC SYSTEM
Type of Device
FLUID JET REMOVAL SYSTEM
Manufacturer (Section D)
PROCEPT BIOROBOTICS CORPORATION
900 island drive
suite 101
redwood city CA 94065 1494
Manufacturer (Section G)
PROCEPT BIOROBOTICS CORPORATION
900 island drive
suite 101
redwood city CA 94065 1494
Manufacturer Contact
doria esquivel
900 island drive
suite 101
redwood city, CA 94065-1494
MDR Report Key16561285
MDR Text Key311695785
Report Number3012977056-2023-00035
Device Sequence Number1
Product Code PZP
UDI-Device IdentifierB614AB20001
UDI-Public+B614AB20001/$$722C01963/16D20220630J
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
DEN170024
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 06/01/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberAB2000
Device Catalogue NumberAB2000
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/16/2023
Initial Date FDA Received03/16/2023
Supplement Dates Manufacturer Received05/17/2023
Supplement Dates FDA Received06/01/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/30/2022
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient SexMale
-
-