MEDTRONIC NEUROMODULATION PERCEPT; STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINS Back to Search Results
Model Number B35200
Device Problems High impedance (1291); Unintended Collision (1429); Application Program Problem (2880)
Patient Problems Scar Tissue (2060); Insufficient Information (4580)
Event Type  malfunction  
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the patient lives in a nursing home.There, the patient had a lot of falls and countless scars on his body.The wife reports that already before christmas there was a worsening of the symptoms.The patient has very poor symptom control on the right side of his body and can no longer touch anything.The wife noticed error code 1703 when reading with the patient's handheld device and informed the physician. the impedances were measured and it was determined that it was an out of regulation (oor) code.The stimulation cannot be delivered because the impedances in the right hemisphere are too high.This affects all four contacts.The session report showed all right stn monopolar contacts were >5k and all bipolar contacts were >10k. the patient immediately went to the emergency room and was cared for by the hcp.It is unknown if the issue resolved.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERCEPT
Type of Device
STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINS
Manufacturer (Section D)
MEDTRONIC NEUROMODULATION
7000 central ave ne
minneapolis MN 55432
Manufacturer (Section G)
MEDTRONIC NEUROMODULATION
7000 central ave ne
minneapolis MN 55432
Manufacturer Contact
glen belmer
7000 central avenue ne rcw215
minneapolis, MN 55432
6122713209
MDR Report Key16564943
MDR Text Key311709137
Report Number2182207-2023-00475
Device Sequence Number1
Product Code MHY
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
P960009
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Consumer,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 03/17/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberB35200
Device Catalogue NumberB35200
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 02/20/2023
Initial Date FDA Received03/17/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-