Model Number 97800
Device Problems High impedance (1291); Application Program Problem (2880); Device Contamination with Chemical or Other Material (2944)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/16/2023
Event Type  malfunction  
Event Description
Information was received from a manufacturer representative regarding a patient who was implanted with an implantable neurostimulator (ins) for urinary dysfunction.It was reported that during the procedure, all of the electrode 1 options were greater than 4000 ohms.The manufacturer representative (rep) and healthcare provider (hcp) disconnected the lead and reconnected and had the same issue.There was blood in the ins header so they stimulated electrode 1 intraoperatively and were able to get a motor response.The hcp opted to finish the surgery.Postoperatively when the rep was programming the patient, all of electrode 0 options were now greater than 4000 ohms and electrode 1 was normal.Program options 1,3,4,5, 7 all maxed out at 0.3 milliamps.The rep ran an impedance check again.They programmed around the 0 electrode that was greater than 4000 ohms and were able to give the patient 5 programs that they felt appropriately.The cause was unknown.The issue was ongoing.
 
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTERSTIM X
Type of Device
STIMULATOR, ELECTRICAL, IMPLANTABLE, FOR INCON
Manufacturer (Section D)
MEDTRONIC NEUROMODULATION
7000 central ave ne
minneapolis MN 55432
Manufacturer (Section G)
MEDTRONIC NEUROMODULATION
7000 central ave ne
minneapolis MN 55432
Manufacturer Contact
glen belmer
7000 central avenue ne rcw215
minneapolis, MN 55432
6122713209
MDR Report Key16572348
MDR Text Key312196968
Report Number2182207-2023-00489
Device Sequence Number1
Product Code EZW
UDI-Device Identifier00763000484668
UDI-Public00763000484668
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970004
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 03/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number97800
Device Catalogue Number97800
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 03/16/2023
Initial Date FDA Received03/20/2023
Date Device Manufactured01/18/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age79 YR
Patient SexMale
-
-