MAUDE Adverse Event Report: SIEMENS HEALTHCARE DIAGNOSTICS INC. RAPIDPOINT 500E BLOOD GAS SYSTEM; ELECTRODE MEASUREMENT, BLOOD-GASES (PCO2, PO2) AND BLOOD PH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIEMENS HEALTHCARE DIAGNOSTICS INC. RAPIDPOINT 500E BLOOD GAS SYSTEM; ELECTRODE MEASUREMENT, BLOOD-GASES (PCO2, PO2) AND BLOOD PH Back to Search Results
Catalog Number 11416755
Device Problem Low Test Results (2458)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/08/2023
Event Type  malfunction  
Event Description
A customer reported discrepant ca++ results on their rp500e when compared to their non-siemens lab instrument.
 
Manufacturer Narrative
Customer stated that the patient was on irenat.Per the instrument manual, "the following guidelines should be followed if a patient is on irenat when ionized calcium is measured: measure ca++ before administering irenat.Do not measure ca++ while the patient is on irenat.Ca++ may be measured 96 hours after the last dose of irenat." the customer has been advised of manual instructions.Instrument operating as intended.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RAPIDPOINT 500E BLOOD GAS SYSTEM
Type of Device
ELECTRODE MEASUREMENT, BLOOD-GASES (PCO2, PO2) AND BLOOD PH
Manufacturer (Section D)
SIEMENS HEALTHCARE DIAGNOSTICS INC.
511 benedict ave.
tarrytown NY 10591
Manufacturer (Section G)
SIEMENS HEALTHCARE DIAGNOSTICS MANUFACTURING LTD.
northern road
chilton industrial estate
sudbury CO10 2XQ
UK   CO10 2XQ
Manufacturer Contact
tom myers
2 edgewater drive
norwood, MA 02062
MDR Report Key16576586
MDR Text Key312210355
Report Number3002637618-2023-00013
Device Sequence Number1
Product Code CHL
Combination Product (y/n)N
Reporter Country CodeFP
PMA/PMN Number
K192240
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Reporter Occupation Physician
Type of Report Initial
Report Date 03/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number11416755
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/10/2023
Initial Date FDA Received03/20/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age54 YR
Patient SexMale
-
-