• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AIZU OLYMPUS CO., LTD. URETERO-RENO FIBERSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AIZU OLYMPUS CO., LTD. URETERO-RENO FIBERSCOPE Back to Search Results
Model Number URF-P7R
Device Problems Device Reprocessing Problem (1091); Improper or Incorrect Procedure or Method (2017)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/27/2023
Event Type  malfunction  
Manufacturer Narrative
Staff were in-serviced on proper pre-cleaning and leak testing steps (found in the olympus manual).The device will not be returned.The investigation is ongoing.A supplemental will be submitted on completion of investigation or if any additional information is available.
 
Event Description
The endoscope specialists (ess) completed a repair reduction observation at the site along with facility summary review.In-service included repair reduction information, tips, techniques, and cost.During observation, facility staff were found not performing any of the pre-cleaning steps at the end of the case using uretero-reno fiberscope.Staff were observed doing leak testing steps out of sequence in the sterile processing department.This report is being submitted to report incorrect reprocessing procedure followed by the site.There was no patient involvement.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the root cause of the incorrect reprocessing was the user¿s understanding differed from olympus recommendation in device handling and/or reprocessing steps.Re-training for the correct handling was completed.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
URETERO-RENO FIBERSCOPE
Type of Device
URETERO-RENO FIBERSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key16577055
MDR Text Key312230124
Report Number9610595-2023-04698
Device Sequence Number1
Product Code FGB
UDI-Device Identifier04953170403835
UDI-Public04953170403835
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K181451
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 05/08/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberURF-P7R
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 02/27/2023
Initial Date FDA Received03/20/2023
Supplement Dates Manufacturer Received04/13/2023
Supplement Dates FDA Received05/08/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/20/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-