(b)(4).The actual device was not returned; however, the customer provided two photos for analysis.The complaint of kit not sealed properly was able to be confirmed by the photos.The images from the customer appear to show a kit with a portion of the lidstock unsealed from the tray flange.However, a complete visual inspection could not be performed the sample was not returned for analysis.A device history record review was performed, and no relevant findings were identified.The instructions for use (ifu) provided with this kit warns the user, "do not use if package is damaged." without the device to evaluate, the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.
|