It was reported to boston scientific corporation that a spyscope ds ii access & delivery catheter was used in the pancreatic duct during an endoscopic retrograde cholangeopancreatography (ercp) and electrohydraulic lithotripsy (ehl) procedure performed on (b)(6) 2023.During the procedure, the image of the spyscope ds ii was lost about two hours of usage.Initially, the ehl probe would not cross and became stuck around the tip of the spyscope ds ii.The image became distorted and eventually lost.The procedure was not completed and reschedule to the next session as the procedure time was prolonged.There were no patient complications reported as a result of this event.
|
It was reported to boston scientific corporation that a spyscope ds ii access & delivery catheter was used in the pancreatic duct during an endoscopic retrograde cholangeopancreatography (ercp) and electrohydraulic lithotripsy (ehl) procedure performed on (b)(6) 2023.During the procedure, the image of the spyscope ds ii was lost about two hours of usage.Initially, the ehl probe would not cross and became stuck around the tip of the spyscope ds ii.The image became distorted and eventually lost.The procedure was not completed and reschedule to the next session as the procedure time was prolonged.There were no patient complications reported as a result of this event.
|
Block h6 (impact codes): imdrf impact code f1001 is being used to capture the reportable event of aborted/cancelled procedure.Block h10: the returned spyscope ds ii was analyzed, and a visual evaluation noted signs of use in the form elevator marks on the shaft of the catheter.Elevator marks were measured approximately from the tip and no damage was noted.Witness marks were observed on the contacts of the umbilicus connector, indicating it was connected to a controller.An image assessment for visualization was performed.The device was plugged into the controller.A live, clear image was displayed.No problems were observed with physical connectivity of the device.The umbilicus connector was visually inspected, and no damage or defects were noted.The device was fully articulated in all directions; no problems were identified with the image.X-ray imaging of the distal tip showed no problems with the redistribution layer (rdl) or thru-silicon vias (tsvs).X-ray imaging of the handle showed no problems with the printed circuit board assembly (pcba) or camera wires.A functional assessment of accessory passability was performed.Testing was conducted with an autolith ehl probe.First, the spyscope was introduced into an articulation fixture to test for passability.No problems were observed and the spyscope passed freely.The ehl probe was passed through the spyscope in a straight configuration and no difficulty was observed.The probe passed freely.The probe was removed from the device with no problem observed.The spyscope device was placed in an articulation fixture and a functional assessment was repeated by passing the probe through the device.The probe passed freely.The probe was removed from the device with no problem observed.The handle was opened and the components within were visually inspected.It was noted that there was procedural residue on the plastic optical fibers (pofs) and camera wire in the breakout region, indicated procedural fluids had flowed back up the optics lumen into the handle during use.The tip was blocked, and saline was flushed through the irrigation tubing to induce backflow of saline into the optics lumen.This caused the image to be disrupted, with blue and orange lines across the screen and a purple hue, leading to a full loss of image.The saline was drained from the optics lumen and the image was restored.The reported event was confirmed.During product analysis, it was noted that procedural residue remained at the top of the optics lumen in the breakout.The optics lumen was filled with saline and the image was disrupted.Draining the optics lumen resulted in the restoration of the image.The image signal does not withstand the change in capacitance created by the introduction of saline into the optics lumen, and procedural factors can cause this fluid to enter the optics lumen during use.An investigation to address this problem has been completed.Based on all gathered information, the probable cause selected for the image failure due to fluid in the optics lumen is cause traced to device design, which indicates that the problems are traced to the design specifications.A labeling review was performed and, from the information available, this device was used per the instructions for use (ifu) / product label.
|