SMITH & NEPHEW, INC. 40MM OXINIUM MODULAR HEAD; PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENTED, METAL/POLYMER, POROUS Back to Search Results
Model Number 71342340
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Hip Fracture (2349)
Event Date 04/19/2021
Event Type  Injury  
Manufacturer Narrative
Internal complaint reference: (b)(4).This complaint was opened by smith+nephew to document a patient complication identified through a review of clinical evidence from a post-market clinical follow-up activity that includes reference to the use of a smith+nephew product.The reported issue(s) relate to known inherent procedural risks that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to confirm a relationship between the reported event and the device or identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are conservatively submitting this report in accordance with applicable regulations.If additional information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.
 
Event Description
On the pmcf titled dca-1000 accord ss clinical dataset, it was reported that, after undergoing right tha, one (1) patient experienced unspecified events that led to a subsequent revision surgery combined with internal fixation on (b)(6) 2021, in which an accord cable system was implanted.The patient suffered an intraoperative acute periprosthetic fracture involving the medial femoral component.The fracture fragment was displaced medially approximately 7 mm.On the lateral view, it was observed that the fracture may extend to the greater trochanteric region.A post-operative follow-up determined a well-healed wound with no evidence of subsidence, loosening, or complicating features.
 
Manufacturer Narrative
Given the nature of the alleged incident, the devices, could not be returned for evaluation.The clinical/medical investigation concluded that, the provided clinical dataset was reviewed; however, as of the date of this medical investigation, patient-specific supporting clinical documentation has not been provided; therefore, there were no clinical factors found which would have contributed to the event.The patient impact beyond the reported events cannot be determined with the information provided.No further clinical assessment is warranted at this time.A review of the production orders did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.For the oxinium modular head, a review of complaint history revealed a similar event for the listed device over the previous 12 months, but no similar events for the batch based on the historical data, this failure mode will be monitored for future complaints for any necessary corrective actions.For the titanium modular head and the redapt stem, a review of complaint history for the part numbers over the past 12 months and for the batch numbers based on historical data of the device did not reveal similar events for the listed devices.A review of the instructions for use documents for total hip systems revealed in warnings and precautions that congenital deformity, improper implant selection, improper broaching or reaming, osteoporosis, bone defects due to misdirected reaming, trauma, improper implant alignment or placement can increase the risk of femoral or pelvic fractures.Also, failure to use the optimum-sized component may result in fracture of the bone.A review of the risk management files revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to these products and event.At this time, we have no reason to suspect that the products failed to meet any specifications at the time of manufacture.Factors that could contribute to the reported event include patient bone quality, patient condition and/or surgical technique.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
40MM OXINIUM MODULAR HEAD
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENTED, METAL/POLYMER, POROUS
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key16583207
MDR Text Key311681694
Report Number1020279-2023-00607
Device Sequence Number1
Product Code MBL
UDI-Device Identifier00885556112410
UDI-Public00885556112410
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K093363
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Study,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 05/30/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number71342340
Device Catalogue Number71342340
Device Lot Number21AM10731
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/16/2023
Initial Date FDA Received03/21/2023
Supplement Dates Manufacturer Received05/26/2023
Supplement Dates FDA Received05/30/2023
Date Device Manufactured01/20/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
RDPT 240MM RV STM SZ14(PN:71354721,LN:16LTM0002F).; TI MDLR HD SLV12/14TPR-4(PN:71344245,LN:19JT27949).
Patient Outcome(s) Required Intervention; Hospitalization;
Patient Age71 YR
Patient SexMale
-
-