It was reported that this right ventricular (rv) lead exhibited pacing lead impedance (pli) measurements that were less than 200 ohms, which was out-of-range (oor).Technical services (ts) analyzed presenting electrogram (egm) and observed that there was loss of capture and no sensing of the rv lead at maximum output as well.It was noted that the patient's intrinsic rhythm was stable.Ts provided troubleshooting and recommended lead revision.It was also noted that there was a break in the insulation of this lead.Isometric testing was performed.Rv sensing was turned off and it was noted that the plan is to extract this lead.No adverse patient effects were reported.Currently, this lead remains in service.
|
It was reported that this right ventricular (rv) lead exhibited pacing lead impedance (pli) measurements that were less than 200 ohms, which was out-of-range (oor).Technical services (ts) analyzed presenting electrogram (egm) and observed that there was loss of capture and no sensing of the rv lead at maximum output as well.It was noted that the patient's intrinsic rhythm was stable.Ts provided troubleshooting and recommended lead revision.It was also noted that there was a break in the insulation of this lead.Isometric testing was performed.Rv sensing was turned off and it was noted that the plan is to extract this lead.No adverse patient effects were reported.Currently, this lead remains in service.According to additional information, this lead was explanted and replaced with a new rv lead.No additional adverse patient effects were reported.This lead will not be returned to boston scientific for investigation.
|