Model Number 550AHCT
Device Problem Electrical Shorting (2926)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The reported complaint was confirmed.During laboratory analysis, the pst observed the power switch connector to be shorted while contacting the removed chassis cover, which caused the connector to be heat damaged and the battery clip to also be heat damaged.He also observed the battery connector to have loose pins.At the time of the event, the alternating current (ac) power was disconnected leaving only the 12 volt (v) back up battery as a power source.It was determined that the power switch connector did not meet specification.The service repair technician replaced the power switch, the battery connector, the printed circuit board assemblies (pcbas), single board computer (sbc) batteries, power supply and associated parts.The product will be brought to manufacturer's specifications before being returned to the customer.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Event Description
The product surveillance technician (pst) reported that during testing of the device at the service center, the blood parameter monitor (bpm) power switch connector shorted.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CDI BLOOD PARAMETER MONITORING SYSTEM 550
Type of Device
CARDIOPULMONARY BYPASS ON-LINE BLOOD GAS
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
6200 jackson road
ann arbor MI 48103
Manufacturer (Section G)
SAME
Manufacturer Contact
douglas patton
6200 jackson road
ann arbor, MI 48103
7346634145
MDR Report Key16584467
MDR Text Key311950033
Report Number1828100-2023-00083
Device Sequence Number1
Product Code DRY
UDI-Device Identifier00886799001790
UDI-Public(01)00886799001790(11)190117
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K182110
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 03/21/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number550AHCT
Device Catalogue Number550AHCT
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 02/24/2023
Initial Date FDA Received03/21/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/17/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-