One infusion pump was received for evaluation.Visual inspection found tamper seal broken, and scratched lens.No evidence in the event history log for the reported issue.Running three accuracy tests, the pump was found to be over delivering to the manufacturing specifications.The reported problem was duplicated.The cause of the reported problem is the expulsor is damaged.The expulsor was replaced.With no indication of a manufacturing defect found during evaluation, no device history review was performed.A review of service history found the device had not previously been in for service.
|