Udi and pma/510k is unknown, no product information has been provided to date.Expiration date and manufacture date are unknown, no information is available based on reported lot number.Month and year of event have been provided ,day is unknown.Five pictures and one sample device were received for investigation.The provided photos demonstrate a torn section on the circuit tubing of the reported device.The reported issue was confirmed in the provided sample device during both visual inspection and functional testing, which identified leakage from a tear in the circuit tubing.No lot number was reported, therefore a review of manufacturing device history records could not be conducted.However the investigation attributed the root cause of this issue to an injection molding fault during the manufacturing process.This issue will continue to be monitored and further actions taken accordingly.
|