A steris service technician inspected the advantage plus aer and found the device to be operating properly.The technician ran a test cycle and found no issues with the function or operation of the advantage plus aer.User facility personnel stated to the steris service technician that prior to and following the reported event, user facility personnel utilized the advantage plus aer with no issues.The employee subject of the reported event was only wearing gloves while operating the advantage plus aer.The operator manual states (pg.85), "warning: avoid possible chemical burns.Always wear personal protective equipment (gloves, goggles) when handling disinfectant and/or detergent." the steris service technician counseled user facility personnel on the proper use and operation of the advantage plus aer specifically, the importance of wearing proper ppe.The steris service technician returned the advantage plus aer to service and no additional issues have been reported.
|