APPLIED MEDICAL RESOURCES C4120, 38CM GRASPER REPOS CART, 10/BX; LAPAROSCOPIC ACCESSORIES, GYNECOLOGIC Back to Search Results
Model Number C4120
Device Problems Material Separation (1562); Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/16/2023
Event Type  malfunction  
Event Description
Procedure performed: lap hernia repair event description: the lattice pad popped off one of the jaws during surgery.The pad was retrieved from the patient.The case was completed with a new device.No patient injury and the patient is doing fine.Product available for return.Intervention: used a new one to complete the case patient status: fine.
 
Manufacturer Narrative
The event device is anticipated to be returned to applied medical for evaluation.A follow-up report will be provided upon completion of investigation.
 
Manufacturer Narrative
The event unit was returned to applied medical for evaluation without a latis pad.Visual inspection confirmed the complainant¿s experience of pad detachment.Based on the condition of the returned unit, the pad detached due to adhesive failure, likely resulting from usage during the procedure.The probability and criticality of harm resulting from this failure have been evaluated and were found to be at an acceptable level.
 
Event Description
Procedure performed: lap hernia repair event description: the lattice pad popped off one of the jaws during surgery.The pad was retrieved from the patient.The case was completed with a new device.No patient injury and the patient is doing fine.Product available for return.Intervention: used a new one to complete the case patient status: fine.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
C4120, 38CM GRASPER REPOS CART, 10/BX
Type of Device
LAPAROSCOPIC ACCESSORIES, GYNECOLOGIC
Manufacturer (Section D)
APPLIED MEDICAL RESOURCES
22872 avenida empresa
rancho santa margarita CA 92688
Manufacturer Contact
farah azmi
22872 avenida empresa
rancho santa margarita, CA 92688
9497138710
MDR Report Key16593165
MDR Text Key312299019
Report Number2027111-2023-00391
Device Sequence Number1
Product Code NWV
UDI-Device Identifier00607915116866
UDI-Public(01)00607915116866(17)260105(30)01(10)1477125
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 07/21/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberC4120
Device Catalogue Number101472471
Device Lot Number1477125
Was Device Available for Evaluation? Device Returned to Manufacturer
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 03/16/2023
Initial Date FDA Received03/22/2023
Supplement Dates Manufacturer Received03/16/2023
Supplement Dates FDA Received07/22/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/06/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-